The Merits of Mechanical CPR

Bonus Videos
 >> Watch how the AutoPulse is applied using two-person pit crew deployment at EMS Today in March 2012.

 Watch how Cypress Creek (Texas) EMS has mastered the rapid application of the LUCUS 2 device.

Approximately 460,000 individuals die every year from out-of-hospital cardiac arrest (OHCA).(1) Studies have consistently shown only an average of 5—15% of patients treated with standard CPR survive cardiac arrest, and providing optimal blood flow to a patient in cardiac arrest by performing quality, uninterrupted CPR is uniformly thought to have a positive impact on improving overall survival.(2,3)

The quality of CPR has been an underappreciated factor and is only now beginning to emerge as an important aspect of successful resuscitation. Manual chest compressions are often done incorrectly, especially in the back of a moving ambulance, and incorrect chest compression can negatively impact survival.(4,5)

Fortunately, there are tools that hold promise for helping providers overcome these inaccuracies: mechanical CPR devices.

The Problem with Manual CPR
The quality of CPR is an important factor that contributes to survival in sudden cardiac arrest. In one experimental study, pigs received 4 minutes of VF and then 9 minutes of CPR before defibrillation. In the group that received suboptimal compressions (1.5 inches depth and 80 per minute), two out of nine pigs survived; in the group that received optimal compressions (2 inches depth and 100 per minute), eight of nine pigs survived.(6)

The study of adult patients from the Resuscitation Outcomes Consortium Cardiac Arrest Epistry with confirmed ventricular fibrillation (VF) or ventricular tachycardia (VT) has shown that increasing chest compression fraction (hands-on time) during out-of-hospital resuscitation of patients with VF/VT is an independent determinant of survival to hospital discharge.(7) However, three human observational studies showed that interruptions of chest compressions were common, averaging 24% to 57% of the total arrest time.(8)

Interrupting CPR can result in precious seconds lost in trying to “reprime” the pump and results in pooling of blood in the right ventricle. In a study looking at this important element of the pathophysiology of cardiac arrest, VF was induced in air-ventilated pigs, after which ventilation was withdrawn. During the first 3 minutes of VF, arterial blood was transported to the venous circulation, with the consequence that the left ventricle emptied and the right ventricle became greatly distended.

It took 2 minutes of mechanical CPR to re-establish an adequate coronary perfusion pressure, which was lost when interrupted. In this pig model, VF caused venous congestion, an empty left heart, and a greatly distended right heart within 3 minutes. Adequate heart massage before and during defibrillation greatly improved the likelihood of return of spontaneous circulation (ROSC).(9)

It is widely accepted that one of the best predictors of ROSC is attaining a coronary perfusion pressure (CPP) of more than 15 mm Hg in animals and humans.(10—18) Every interruption of CPR causes the CPP to drop immediately.In the same pig study referenced above, when VF was induced, CPP fell from 60 mmHg to 15 mmHg in 15 seconds and continued to plummet into negative numbers. After CPR was restarted, the CPP remained negative for the first minute; it took 90 seconds for it to get back up to 15 mmHg.

Another pig study demonstrated that mean CPP was 20 mmHg in the mechanical group compared to around 5 mmHg in the manual group. All eight pigs in the mechanical group achieved ROSC, as compared with just three pigs in the manual group.(19)

With manual CPR, many factors come into play, including fatigue, physical abilities, focus on several simultaneous tasks, poor-quality CPR during transportation on a stretcher and in the back of a transport vehicle, interruptions during movement of patient and variations in technique and training.

Just maintaining the proper rate is a major challenge. In one study, it was demonstrated that manual CPR resulted in a rate of less than 80 compressions/minute 37% of the time.(20) Other studies looked at the challenges in performing correct CPR in various settings. They found the following percentages of correctly performed CPR:
 54—78% on the floor;
“¢ 46% in the back of a moving ambulance; and
“¢ 21% while transporting on a stretcher.(21,22)

Besides the poor-quality CPR that occurs in the back of a moving ambulance, the risk of injury to the crew members is unacceptable. It has been reported that ambulance personnel who perform CPR in a moving ambulance are at least four times more like to have a fatal or incapacitating injury than personnel who are restrained.

Mechanical CPR allows high-quality CPR to be performed while EMS personnel are restrained; this reduces risk of injury to the crew.(23,24) Restrained ambulance occupants involved in a crash have been shown to have 3.77 times lower risk of fatality and 6.49 times lower risk of incapacitating injury than unrestrained occupants.(24)

Consistent-Quality CPR
The use of automatic mechanical CPR devices can provide a consistent rate and depth of compression and is one way of improving the quality of chest compression and contributing to EMS personnel safety.

There is compelling physiological and animal data suggesting that mechanical chest-compression devices are more effective than manual CPR. At best, standard manual CPR produces coronary and cerebral perfusion that is just 30% of normal.(25)

Some studies that demonstrate the advantages of mechanical CPR devices include:
“¢ Increased brain flow by as much as 60% vs. manual measured in pigs.(26,27)
“¢ Adequate heart pressures–exceeding 15 mmHg in pigs19 and humans–significantly improved over manual CPR.(28,29)
“¢ Increased EtCO2 levels in prehospital cluster randomized patients.(30)
“¢ Increased hands-on times to up to 90% of the time–significantly improved over manual.(31)

A study from Kramer-Johansen et al states, “The poorer results with sinusoidal than trapezoidal chest compressions add to the arguments that mechanical chest compressions can ensure better quality CPR for more than short intervals of time, unless the logistics around applying a mechanical device has too many negative effects.”(23)

Another advantage of mechanical CPR: it can allow for defibrillation during compressions. In a study on the LUCAS device, it showed that defibrillations did not compromise CPR and can be delivered during ongoing mechanical compressions without any pauses.(32)

These devices are also helpful in maintaining perfusion during cardiac arrest while undergoing a cardiac catheterization interventional procedure. The AHA 2010 Guidelines (Part 12: Cardiac Arrest in Special Situations) state: “Mechanical chest compression devices have been used successfully in an animal model and adult humans to provide maintenance of circulation in cardiac arrest while continuing a percutaneous coronary procedure. It is reasonable to use mechanical CPR during PCI (Class IIa, LOE C).”(33)

Mechanical CPR Concerns
Mechanical CPR is a recent development and as such, the literature around its effect on outcomes is still evolving. Not everyone is a fan of mechanical CPR. A recent systematic literature review of mechanical CPR devices compared to manual CPR during OHCA and ambulance transport concluded that there was insufficient evidence to support or refute the use of mechanical CPR devices in these settings. The study also said that while there was some low-quality evidence suggesting that mechanical CPR can improve consistency and reduce interruptions in chest compressions, there is no evidence that mechanical CPR devices improve survival and may worsen neurological outcome.(34)

One area of concern with mechanical CPR devices has been the time off the chest while applying the device. However, in our experience with the LUCAS device in Columbus, Ohio, we have found that application takes only 15—60 seconds and that the time off the chest is minimal. In addition, it is important to take into account the total time on and off the chest. In a prehospital real-use study, the hands-off ratio with manual CPR was on average 22%; however, with mechanical CPR it was just 9%, as measured over the entire resuscitation episodes of approximately 33—40 minutes.

The interruptions in manual CPR increased from 19% on the scene to 27% during transport, whereas those with the mechanical CPR device stayed at the same low ratio (8—10%) both on scene and during transport.(34)

Another more practical concern is the cost of these devices. They can cost nearly the same as a monitor/defibrillator, a cost that many feel is too high for a mechanical device that is not as technology-packed as a cardiac monitor. In addition, one unit sold in the U.S., the ZOLL AutoPulse, has a one-time use band that must be replaced after each use; and this replacement cost is not currently reimbursed by most insurance agencies.

Types of Mechanical CPR Devices
There are several types of mechanical CPR devices.

Piston CPR Devices: Piston compression devices use battery power (e.g., the LUCAS 2 from Physio-Control) or pneumatic power derived from compressed oxygen (e.g., the LifeStat from Michigan Instruments, which also features an associated integrated ventilator). Early studies found that mechanical piston CPR improved hemodynamic parameters, but not survival.(35—38) However, these studies were performed prior to the latest AHA guidelines and before crews began using compressions-only CPR, early adult IO (EZ-IO) establishment, impedance threshold devices (ResQPOD), therapeutic hypothermia and a “pit crew” approach to resuscitation.

Piston chest compression devices have been shown to be actually less damaging than manual CPR.(39,40) Trauma, such as rib fractures, is an occasional occurrence in CPR and a complication of this and other mechanical devices.

The 2010 AHA Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care Science state, “There is insufficient evidence to support or refute the routine use of mechanical piston devices in the treatment of cardiac arrest. Mechanical piston devices may be considered for use by properly trained personnel in specific settings for the treatment of adult cardiac arrest in circumstances (e.g., during diagnostic and interventional procedures) that make manual resuscitation difficult (Class IIb, LOE C).”(41)

However, EMS insiders believe that the positive results being realized by EMS systems using mechanical CPR devices applied early and efficiently; the move to compressiond-only CPR and its proven efficacy; and the results of Dr. Idris’ study (see p. 4 of this supplement), which shows the need to keep compressions within a defined range of 100—120 compressions per minute, will result in a stronger position for mechanical CPR in the future.

Active Compression-Decompression (ACD) Devices: This piston technique, not yet approved for use in the U.S., integrates a suction cup that allows for active return of the chest and provides both active chest compression and active chest decompression (ACD-CPR).

A 2005 study demonstrated that incomplete chest recoil during CPR worsens hemodynamics. The study, although performed in piglets, indicated that leaning on the chest during CPR and not allowing the chest to fully expand on the upstroke led to worse systolic blood pressure, worse coronary perfusion pressure and substantially reduced myocardial blood flow.(42)

Load-Distributing Band (LDB): The ZOLL AutoPulse device uses a load-distributing band that is placed around the patient’s chest and circumferentially tightens and loosens around the patient, distributing the force evenly to generate changes in intrathoracic pressure.

Initial studies of the device had found improved hemodynamic parameters and coronary perfusion pressures as well as improved prehospital survival to the emergency department.(43—47) Recent trials showed improved hemodynamics, with coronary perfusion pressures above the level generally associated with improved survival, as well as improvement in survival to arrival at the emergency department when compared with manual CPR.(17,43)

A multi-site cluster-randomized trial (ASPIRE) that the Columbus (Ohio) Division of Fire participated in was terminated early due to a lack of benefit and apparent harm.(48) It is believed that site-specific factors had led to the poorer results,49 and further clinical research is still ongoing.

However, in the 2006 JAMA article, the authors looked at 783 adults with non-traumatic cardiac arrest in the prehospital setting.(49) They compared resuscitation outcomes before and after switching over their urban EMS system to a load-distributing band (LDB) CPR device. They compared 499 patients in the manual CPR phase and 284 patients in the LDB CPR phase. The LDB device was placed on 210 out of the 284 patients in this last phase.

Rates for ROSC and survival were increased with LDB CPR compared with manual CPR. Survival to hospital admission was 20.9% in the LDB CPR group vs. 11.1% in the manual CPR group, and survival to hospital discharge was 9.7% vs 2.9% respectively.

In secondary analysis of the 210 patients on whom the LDB device was applied, 38 patients (18.1%) survived to hospital admission (95% CI, 13.4%—23.9%) and 12 patients (5.7%) survived to hospital discharge (95% CI, 3.0%—9.3%). Among patients in the manual CPR and LDB CPR groups who survived to hospital discharge, there was no significant difference between groups in cerebral performance category (P=.36) or overall performance category (P=.40).

Their final conclusion: Compared with resuscitation using manual CPR, a resuscitation strategy using LDB CPR on EMS ambulances is associated with improved survival to hospital discharge in adults with out-of-hospital nontraumatic cardiac arrest.

According to the 2010 AHA Guidelines, “The LDB may be considered for use by properly trained personnel in specific settings for the treatment of cardiac arrest (Class IIb, LOE B). However, there is insufficient evidence to support the routine use of the LDB in the treatment of cardiac arrest.”(41)

The Columbus Experience
The Columbus (Ohio) Division of Fire has had the opportunity to use and evaluate several types of mechanical CPR devices. We used the Thumper devices in the 1970s and 1980s, and we were one of the participating systems in the ASPIRE trial, which provided exposure to the Autopulse device.

In July 2011, we put LUCAS devices on three of our EMS officer vehicles. In the first six months of usage, we used the devices 71 times. Although this was not a scientific study, our average ROSC rate with manual CPR was 30.5% and our hospital discharge rate was just 11.1%. With the LUCAS device it was 19% and 2.3%.

During the first six months of 2012, the ROSC rate and hospital discharge rate with manual CPR was 33% and 12% respectively and with the LUCAS it was 30% and 4%.

Our results are not as positive as some other systems that have mechanical CPR devices on all of their ALS vehicles because the LUCAS devices were only available on our EMS officer vehicles. Although manual CPR had been going on from the start of each arrest, the device was being applied late into most cardiac arrests. It was also being applied to patients who had not quickly converted from a ventricular fibrillation or ventricular tachycardia rhythm.

The Columbus Department of Fire EMS staff has a favorable attitude about mechanical CPR devices. The advantages in terms of job enhancement have already been discussed. In a survey we conducted, our EMS personnel reported the following:

“¢ 60% of providers surveyed said application time for the LUCAS device was less than 30 seconds; 40% said it was 30—60 seconds.
“¢ 100% of providers surveyed said the LUCAS 2 device makes the treatment of cardiac arrest patients in the field go smoother than with manual CPR.
“¢ 100% of providers surveyed said they perceived that they obtain ROSC (pulses) more often on patients that have the LUCAS device applied than with manual CPR.
“¢ 100% of providers surveyed said they see no drawbacks to using mechanical CPR in the field.
“¢ Responders identified the following advantages to using the LUCAS device in the field: It gives them more time to focus on other skills and patient care; it is safer for transport in the back of the medic (no one standing to perform CPR); it can be applied rapidly and, once applied, provides consistent CPR with no interruptions.

Summary
In the Cochrane review, a large literature review of randomized controlled trials (RCTs), cluster RCTs and quasi-randomized studies that compared mechanical chest compressions to manual chest compressions, the authors conclude: “There is insufficient evidence from human RCTs to conclude that mechanical chest compressions during cardiopulmonary resuscitation for cardiac arrest are associated with benefit or harm. Widespread use of mechanical devices for chest compressions during cardiac is not supported by this review. More RCTs that measure and account for CPR process in both arms are needed to clarify the potential benefit from this intervention.”(50)

Further studies are needed to truly determine the efficacy of mechanical CPR devices. The LINC study, a large RCT comparing LUCAS to manual CPR, is ongoing and expected to complete enrollment in the next few months. The LINC trial is a prospective randomized multicenter study where LUCAS will be used with simultaneous defibrillation compared to manual chest compressions. The study is being conducted in first-tier systems, where LUCAS always will be placed in the first-arriving ambulance.

There also has been promising research showing benefit of Active Compression Decompression CPR and it is expected that the LUCAS device and others will utilize this method of CPR in the future. Studies with these devices will also be critical in assessing the overall survival of patients receiving combined mechanical and ACD CPR.

David P. Keseg, MD, FACEP, is the medical director for the Columbus (Ohio) Division of Fire and associate professor of emergency medicine at the Ohio State University Wexner Medical Center. He can be reached at dkeseg@columbus.gov.

Disclosure: The author has reported no conflicts of interest with the sponsors of this supplement.

References
1. Zheng ZJ, Croft JB, Giles WH, et al. Sudden cardiac death in the United States, 1989 to 1998. Circulation.2001;104(18):2158—2163.
2. Tunstall-Pedoe H, Bailey L, Chamberlain DA, et al. Survey of 3765 cardiopulmonary resuscitations in British hospitals (the BRESUS Study): methods and overall results. BMJ. 1992;304(6838):1347—1351.
3. Halperin HR & Weisfeldt ML. New approaches to CPR. Four hands, a plunger, or a vest. JAMA. 1992;267(21):2940—2941.
4. Ochoa FJ, Ramalle-Gomara E, Lisa V, et al. The effect of rescuer fatigue on the quality of chest compressions. Resuscitation. 1998;37(3):149—52.
5. Van Hoeyweghen RJ, Bossaert LL, Mullie A, et al. Quality and efficiency of bystander CPR. Belgian Cerebral Resuscitation Study Group. Resuscitation. 1993;26(1):47—52.
6. Wu JY, Li CS, Liu ZX, et al. A comparison of 2 types of chest compressions in a porcine model of cardiac arrest. Am J Emerg Med. 2009;27(7):823—829.
7. Christenson J, Andruskiek D, Everson-Stewart S, et al. Chest compression fraction determines survival in patients with out-of-hospital ventricular fibrillation. Circulation. 2009; 120(13):1241—1247.
8. Berg RA, Hemphill R, Abella BS, et al. Part 5: Adult Basic Life Support: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2010;122(18_suppl_3):S690.
9. Steen S, Liao Q, Pierre L, et al. The critical importance of minimal delay between chest compressions and subsequent defibrillation: a haemodynamic explanation. Resuscitation. 2003;8(3):249—258.
10. Halperin HR, Guerci AD, Chandra N, et al. Vest inflation without simultaneous ventilation during cardiac arrest in dogs: improved survival from prolonged cardiopulmonary resuscitation. Circulation. 1986;74(6)1407—1415.
11. Ralston SH, Voorhees WD, Babbs CF. Intrapulmonary epinephrine during prolonged cardiopulmonary resuscitation: improved regional blood flow and resuscitation in dogs. Ann Emerg Med. 1984;13(2):79—86.
12. Michael JR, Guerci AD, Koehler RC, et al. Mechanisms by which epinephrine augments cerebral and myocardial perfusion during cardiopulmonary resuscitation in dogs. Circulation. 1984;69(4):822—835.
13. Kern KB, Ewy GA, Voorhees WD, et al. Myocardial perfusion pressure: a predictor of 24-hour survival during prolonged cardiac arrest in dogs. Resuscitation. 1988;16(4):241—250.
14. Sanders AB, Ewy GA, Taft TV. Prognostic and therapeutic importance of the aortic diastolic pressure in resuscitation from cardiac arrest. Crit Care Med. 1984;(12)871—873.
15. Sanders AB, Ogle M, Ewy GA. Coronary perfusion pressure during cardiopulmonary resuscitation. Am J Emerg Med. 1985;3(1)11—14.
16. Wolfe JA, Maier GW, Newton JR Jr, et al. Physiologic determinants of coronary blood flow during external cardiac massage. J Thorac Cardiovasc Surg. 1988;95(3):523—532.
17. Paradis NA, Martin GB, Rivers EP, et al. Coronary perfusion pressure and the return of spontaneous circulation in human cardiopulmonary resuscitation. JAMA. 1990;263(8):1106—1113.
18. McDonald JL. Coronary perfusion pressure during CPR in human beings [abstract]. Ann Emerg Med. 1983;12:144.
19. Liao Q, Sjà¶berg T, Paskevicius A, et al. Manual versus mechanical cardiopulmonary resuscitation. An experimental study in pigs. BMC Cardiovasc Disord. 2010;10:53.
20. Abella BS, Sandbo N, Vassilatos P, et al. Chest compression rates during cardiopulmonary resuscitation are suboptimal: a prospective study during in-hospital cardiac arrest. Circulation. 2005;111(4):428—434.
21. Stone CK & Thomas SH. Can correct closed-chest compressions be performed during prehospital transport? Prehosp Disaster Med. 1995;10(2):121—123.
22. Kim JA, Vogel D, Guimond G, et al. A randomized, controlled comparison of cardiopulmonary resuscitation performed on the floor and on a moving ambulance stretcher. Prehosp Emerg Care. 2006;10(1):68—70.
23. Kramer-Johansen J, Pytte M, Tomlinson AE et al. Mechanical chest compressions with trapezoidal waveform improve haemodynamics during cardiac arrest. Resuscitation. 2011;82(2):213—218.
24. Becker L, Zaloshnja E, Levick N, et al. Relative risk of injury and death in ambulances and other emergency vehicles. Accid Anal Prev. 2003;35(6):941—948.
25. Nolan JP, Soar J, Zideman D, et al. European Resuscitation Council Guidelines for Resuscitation 2010. Section 1. Executive Summary: CPR techniques and devices. Resuscitation. 2010;81(10):1219—1276.
26. Rubertsson S & Karlsten R. Increased cortical cerebral blood flow with LUCAS; a new device for mechanical chest compressions compared to standard external compressions during experimental cardiopulmonary resuscitation. Resuscitation. 2005;65(3):357—363.
27. Jimenez FC, Padro PP, Garcia AS, et al. Cerebral flow improvement during CPR with LUCAS, measured by Doppler. [abstract] Resuscitation. 2011;82S1:S1—S34.
28. Wagner H, Madsen Hardig B, Harnek J, et al. Aspects on resuscitation in the coronary interventional catheter laboratory. Circulation. 2010;122:A91.
29. Larsen AI, Hjà¸rnevik A, Bonarjee V, et al. Coronary blood flow and perfusion pressure during coronary angiography in patients with ongoing mechanical chest compression: a report on 6 cases. Resuscitation. 2010;81(4):493—497.
30. Axelsson C, Karlsson T, Axelsson AB, et al. Mechanical active compression-decompression cardiopulmonary resuscitation (ACD-CPR) versus manual CPR according to pressure of end tidal carbon dioxide (P(ET)CO2) during CPR in out-of-hospital cardiac arrest (OHCA). Resuscitation. 2009;80(10):1099—1103.
31. Olasveengen TM, Wik L, Steen PA. Quality of cardiopulmonary resuscitation before and during transport in out-of-hospital cardiac arrest. Resuscitation. 2008;76(2):185—190.
32. Tabereaux PB, Walcott GP, Rogers JM, et al. Activation patterns of Purkinje fibers during long-duration ventricular fibrillation in an isolated canine heart model. Circulation. 2007;116(10):1113—1119.
33. Vanden Hoek TL, Morrison LJ, Shuster M, et al. Part 12: Cardiac Arrest in Special Situations: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2010;122(18_suppl_3):S849.
34. Ong EH, Mackey KE, Zhang ZC, et al. Mechanical CPR devices compared to manual CPR during out-of-hospital cardiac arrest and ambulance transport: a systematic review. Scand J Trauma Resusc Emerg Med. 2012;20(1):39.
35. McDonald JL. Systolic and mean arterial pressures during manual and mechanical CPR in humans. Ann Emerg Med. 1982; 11(6):292—295.
36. Ward KR, Menegazzi JJ, Zelenak RR, et al. A comparison of chest compressions between mechanical and manual CPR by monitoring end-tidal PCO2 during human cardiac arrest. Ann Emerg Med. 1993;22(4):669—674.
37. Dickinson ET, Verdile VP, Schneider RM, et al. Effectiveness of mechanical versus manual chest compressions in out-of-hospital cardiac arrest resuscitation: a pilot study. Am J Emerg Med. 1998;16(3):289—292.
38. Taylor GJ, Rubin R, Tucker M, et al. External cardiac compression. A randomized comparison of mechanical and manual techniques. JAMA. 1978;240(7):644—646.
39. Wik L. Automatic and manual mechanical external chest compression devices for cardiopulmonary resuscitation. Resuscitation. 2000;47(1):7—25.
40. Allen JR. The use of life aid cardiopulmonary resuscitator–Preliminary report. Br J Clin Pract. 1974;28(8):286—288.
41. Cave DM, Gazmuri RJ, Otto CW, et al. Part 7: CPR Techniques and Devices: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2010;122(18_suppl_3):S723.
42. Yannopoulos D, McKnite S, Aufderheide TP, et al. Effects of incomplete chest wall decompression during cardiopulmonary resuscitation on coronary and cerebral perfusion pressures in a porcine model of cardiac arrest. Resuscitation. 2005;64(3):363—72.
43. Timerman S, Cardoso LF, Ramires JA, et al. Improved hemodynamic performance with a novel chest compression device during treatment of in-hospital cardiac arrest. Resuscitation. 2004;61(3):273—280.
44. Halperin HR, Paradis N, Ornato JP, et al. Cardiopulmonary resuscitation with a novel chest compression device in a porcine model of cardiac arrest: improved hemodynamics and mechanisms. J Am Coll Cardiol. 2004; 44(11):2214—2220.
45. Halperin H, Berger R, Chandra N, et al. Cardiopulmonary resuscitation with a hydraulic-pneumatic band. Crit Care Med. 2000;28(11 suppl):N203—206.
46. Duchateau FX, Gueye P, Curac S, et al. Effect of the AutoPulse automated band chest compression device on hemodynamics in out-of-hospital cardiac arrest resuscitation. Intensive Care Med. 2010;36(7):1256—1260.
47. Casner M, Andersen D & Isaacs SM. The impact of a new CPR assist device on rate of return of spontaneous circulation in out-of-hospital cardiac arrest. Prehosp Emerg Care. 2005;9(1):61—67.
48. Hallstrom A, Rea TD, Sayer MR, et al. Manual chest compression vs. use of an automated chest compression device during resuscitation following out-of-hospital cardiac arrest: A randomized trial. JAMA. 2006;295(22):2620—2628.
49. Ong ME, Ornato JP, Edwards DP, et al. Use of an automated, load-distributing band chest compression device for out-of-hospital cardiac arrest resuscitation. JAMA. 2006;295(22):2629—2637.
50.Brooks SC, Bigham BL, Morrison LJ. Mechanical versus manual chest compressions for cardiac arrest. Cochrane Database Syst Rev. 2011;19(1):CD007260.