Caution Urged on Powerheart and CardioVive Automated External Defibrillators


 
 

U.S. Fire Administration | | Friday, February 5, 2010


Caution Urged on Powerheart and CardioVive Automated External Defibrillators

The Food and Drug Administration (FDA) is alerting users of certain Powerheart and CardioVive automated external defibrillators that these devices may malfunction unexpectedly. This can cause a failure to deliver the electric shocks that are needed to treat sudden cardiac arrest or arrhythmias.

FDA is alerting users of certain Powerheart and CardioVive automated external defibrillators that these devices may malfunction unexpectedly. This can cause a failure to deliver the electric shocks that are needed to treat sudden cardiac arrest or arrhythmias.



The manufacturer, Cardiac Science Corporation, has received multiple complaints that these devices have malfunctioned, and that their self-test features may not show that there's anything wrong. Before each use, operators are supposed to check for a green light on the front of the AED that shows the device is rescue-ready. But the light may come on even when certain electronic components inside the device have failed.



Malfunctions could cause these AEDs to interrupt ECG analysis and thus prevent shock delivery, or fail to recognize that the pads have been placed. Malfunctions could also cause interference or background noise, which could prevent the device from accurately analyzing heart rhythm and delivering a shock.



About 300,000 Cardiac Science G3 series Powerheart and CardioVive AEDs are affected by this problem worldwide. They were manufactured between August 2003 and August 2009. FDA recommends that users contact Cardiac Science Corporation immediately to arrange for repairs or replacement.



Until the company repairs or replaces the affected devices, FDA recommends that users switch to alternate AEDs if they are available, or that they consider obtaining another AED. If an alternate AED is not available, trained responders should use manual defibrillators. If a manual defibrillator is not available, responders should use the Powerheart or CardioVive AEDs, since they may still be able to deliver the necessary shock.



In the home or in public places, FDA recommends continuing to use these devices if an alternate AED is not available. That's because the consequences of not trying to defibrillate a patient outweigh the risk that the AED may fail.



FDA is gathering more data about this situation, and as we learn more, we will share any new information that might affect the use of these AEDs.

Additional Information:




Connect: Have a thought or feedback about this? Add your comment now
Related Topics: Industry News

What's Your Take? Comment Now ...

Featured Careers & Jobs in EMS

 

 

Get JEMS in Your Inbox

 

Fire EMS Blogs


Blogger Browser

Today's Featured Posts

 

EMS Airway Clinic

Innovation & Progress

Follow in the footsteps of these inspirational leaders of EMS.
More >

Multimedia Thumb

Abilene Loses Helicopter Service

Native Air leaves city with only one air helicopter service.
Watch It >


Multimedia Thumb

D.C. Fire Chief Proposes another Controversial Ambulance Plan

Staffing change will leave immediate neighborhood without fire apparatus.
Watch It >


Multimedia Thumb

FDIC 2014 CHAT: MIKE MCEVOY AND A.J. HEIGHTMAN

Mike McEvoy and A.J. Heightman discuss some new EMS technology at FDIC 2014.
Watch It >


Multimedia Thumb

D.C. Lieutenant in Patient Death May Go Unpunished

Family upset that officer in charge may retire without any discipline.
Watch It >


Multimedia Thumb

The AmbuBus®, Bus Stretcher Conversion Kit - EMS Today 2013

AmbuBus®, Bus Stretcher all-hazards preparedness & response tool
Watch It >


Multimedia Thumb

Field Bridge Xpress ePCR on iPad, Android, Kindle Fire

Sneak peek of customizable run forms & more.
Watch It >


Multimedia Thumb

Braun Ambulances' EZ Door Forward

Helps to create a safer ambulance module.
Watch It >


More Product Videos >