Determining the Effects of Nail Polish in Pulse Oximetry Readings

Review of: Hinkelbein J, Genzwuerker HV, Sogl R, et al: "Effect of nail polish on oxygen saturation determined by pulse oximetry in critically ill patients." Resuscitation. 72(1):82-91, 2007.

The Science

Pulse oximetry has been around for many years, and when they first came out, we were all taught to remove the patient's nail polish to improve the accuracy of the reading. The authors of this study undertook an experiment to determine whether or not the presence of nail polish affects the reading.

They applied nine shades of nail polish to the fingers of patients in the critical care unit who also had arterial lines established. They then compared the oxygen saturation level with that obtained from direct analysis of arterial blood samples.

They found that black, purple, and dark blue resulted in the greatest deviation from the arterial measurement, but it was less than 2%, which is within the margin of error for the device as stated by the manufacturer. If the probe was rotated 90 degrees and applied side-ways on the finger, the effect of the darker colors was eliminated.

When there was a discrepancy between the pulse oximeter and measured levels, the pulse oximeter tended to underestimate the oxygen saturation.

Their conclusion was that the presence of nail polish did not significantly affect the reading, and that routine removal of nail polish is not necessary.

The Street

It is amazing that there are so many things we do in EMS that we simply have not applied any type of scientific scrutiny to. We have all had the unpleasant experience of opening nail polish removal packets to get rid of every bit of polish, only to find the reading is normal.

According to these researchers, this practice is not necessary. However, there are some limitations to this study that needed to be examined. All of their patients were white, normotensive and not hypoxic. They only tested one model of oximeter, the Seimans Nellcor device that emits two wavelengths of light. What, if any, effect these variables would have on the result are not known and should prompt additional research.

Before reading this study, I had never heard of rotating the probe to get a better reading. Those that have done this tell me that sometimes they don't get a good pulse wave indication with normal positioning, but that if the probe is rotated, the pulse wave improves and provides a reading.

RELATED ARTICLES

Rethink the Way EMS Does Patient Care Reports

EMS must let go of the idea prehospital care documentation is strictly about billing and creating a legally defensible record of events.

Hands On May 2015

Read about the latest products in EMS.

Study Examines EMS' Administration of Aspirin

Less than half of eligible patients receive the drug.

Early Clinical Trials Suggests Hydroxocobalamin Beneficial for Hemorrhagic Shock

Medically facilitated hemorrhage control study yields good results, more questions.

Crew Resource Management Can Improve Crew Efficiency

Crews performing in unison offer better learning opportunities.

Serving the Psychological Needs of Your Employees

How does your agency help employees cope with the traumas and stressors of EMS?

Features by Topic

Featured Careers

 

JEMS TV

FEATURED VIDEO TOPICS

Learn about new products and innovations featured at EMS Today 2015

 

JEMS Connect

CURRENT DISCUSSIONS

 
 

EMS BLOGS

Blogger Browser

Today's Featured Posts