Defibtech Announces AED Battery Pack Recall

In rare instances, when the AED is used with an affected battery pack, the AED may falsely detect an error condition during charging for a shock.


 
 

U.S. Fire Administration | | Thursday, June 10, 2010


Defibtech, LLC, is initiating a voluntary recall of 5,418 DBP-2800 Battery Packs used in the Lifeline AED® and ReviveR AEDTM (semi-automatic external defibrillators). This recall affects all DBP- 2800 Battery Packs shipped prior to June 4, 2007. In rare instances, when the AED is used with an affected battery pack, the AED may falsely detect an error condition during charging for a shock, then cancel charge and not provide therapy. Defibtech determined the need for this recall after learning of four reports from end users of this malfunction during patient use.

The company has identified recommendations for the end customer to follow until the battery pack has been corrected, which allows the battery pack to remain in service. A copy of these recommendations is being mailed to all affected customers. This customer notification, as well as instructions on determining whether a battery pack is affected, can also be found on the www.defibtech.com/batteryFA1 web page. For additional information regarding this recall, please refer to the above referenced web page, contact your distributor, or contact Defibtech at techsupport@defibtech.com, 1-877-453-4507 or 1- 203-453-4507.

Defibtech will provide customers with a free battery pack update card to address this issue for all affected battery packs. The correction to the battery pack will be able to be performed at the location where the battery pack is deployed using any DDU-100 series AED and a Defibtech supplied battery pack update card. The battery pack update is expected to be available within the next two weeks.

The DBP-2800 battery packs affected by this recall have been distributed globally to fire departments, EMS, health clubs, schools, and other organizations. The affected battery packs are used in AEDs which can be identified by the words "Lifeline AED®" and "ReviveR AEDTM" on the front of the device.

The Food and Drug Administration (FDA) has determined that this action is a Class I recall. Any adverse reactions experienced with the use of this product and/or quality problems should also be reported to the FDA’s MedWatch Program by phone at 1-800- FDA-1088, or on the MedWatch website at www.fda.gov/medwatch2.




Connect: Have a thought or feedback about this? Add your comment now
Related Topics: News, AED

What's Your Take? Comment Now ...

Featured Careers & Jobs in EMS

 

 

 

 

 

 

 

 

 

Get JEMS in Your Inbox

 

Fire EMS Blogs


Blogger Browser

Today's Featured Posts

 

EMS Airway Clinic

The Evolution of Civilian High Threat Medical Guidelines

How mass killing events have proven a need for new guidelines.
More >

Multimedia Thumb

Ottawa Shooting, EMS Audio

Edited audio from the Parliament Hill incident.
Watch It >


Multimedia Thumb

Illinois Ambulance Service on Chopping Block

Peoria Heights EMS depends on property tax vote.
Watch It >


Multimedia Thumb

Kentucky Firefighters, Medics Drill for Ebola

Lexington firefighters and medics prepare for possible patients.
Watch It >


Multimedia Thumb

Mid-South EMTs Prepare for Ebola

Mid-South EMTs are certified for service.
Watch It >


Multimedia Thumb

Ottawa Shooting Incident

Solider is shot by gunman at national memorial in Ottawa.
More >


Multimedia Thumb

Life Link III Trauma Tactics Conference in Minnesota

Conference was designed to enhance the skills of providers of all levels, covering rescue and prehospital situations, to transport and in-hospital treatment.
More >


Multimedia Thumb

EMS Tailgating

Rigs converted for football.
More >


Multimedia Thumb

VividTrac offered by Vivid Medical - EMS Today 2013

VividTrac, affordable high performance video intubation device.
Watch It >


Multimedia Thumb

The AmbuBus®, Bus Stretcher Conversion Kit - EMS Today 2013

AmbuBus®, Bus Stretcher all-hazards preparedness & response tool
Watch It >


Multimedia Thumb

Field Bridge Xpress ePCR on iPad, Android, Kindle Fire

Sneak peek of customizable run forms & more.
Watch It >


Multimedia Thumb

Braun Ambulances' EZ Door Forward

Helps to create a safer ambulance module.
Watch It >


More Product Videos >