FDA to Review Some External Biphasic Defibrillators' Energy Levels - @ JEMS.com


FDA to Review Some External Biphasic Defibrillators' Energy Levels


 
 

U.S. Food and Drug Administration | | Wednesday, November 11, 2009


FDA notified healthcare professionals that it is investigating energy levels in external biphasic defibrillators with shocks of less than or equal to 200 Joules (J). FDA received reports of 14 events since 2006 in which a 200 J biphasic defibrillator was ineffective in providing defibrillation/cardioversion therapy to a patient, whereas a subsequent shock from a different 360 J biphasic defibrillator resulted in immediate defibrillation/cardioversion. The majority of events occurred during attempts at cardioversion of atrial fibrillation, but there was at least one instance with defibrillation of a ventricular arrhythmia as well. Analysis of the 14 cases does not suggest the need for any change to current clinical practice, and as FDA continues its evaluation of this situation, providers are encouraged to follow the American Heart Association's guidelines/algorithms for treatment of cardiac arrhythmias, and to follow manufacturers' instructions for using defibrillators.

FDA is seeking additional information in order to interpret the significance of these events, and to determine whether FDA activities are advised. If you suspect a problem with a defibrillator, such as a situation where a patient received shocks from multiple devices, the problem should be reported to the FDA'sMedWatch Safety Information and Adverse Event Reporting Program online (atwww.fda.gov/MedWatch/report.htm,) by phone 800/332-1088, or by returning the postage-paid FDA form 3500 (which may be downloaded from the MedWatch"Download Forms" page) by mail (to address on the pre-addressed form) or fax (800/FDA-0178.)

[11/05/2009 --http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm189259.htm-- FDA]




Connect: Have a thought or feedback about this? Add your comment now
Related Topics: Leadership and Professionalism, Cardiac and Circulation

What's Your Take? Comment Now ...

Buyer's Guide Featured Companies

Featured Careers & Jobs in EMS

Get JEMS in Your Inbox

 

Fire EMS Blogs


Blogger Browser

Today's Featured Posts

 

EMS Airway Clinic

Improving Survival from Cardiac Arrest Using ACD-CPR + ITD

Using active compression-decompression CPR with an ITD has been shown to improve 1-year survival from cardiac arrest by 33%.
More >

Multimedia Thumb

Philadelphia Fire Department Apologizes for Medic’s Jab at Police

Union head calls photos a slap in the face of officers.
Watch It >


Multimedia Thumb

D.C. Fire and EMS Crews Blame New Technology for Patient’s Death

Delayed response blamed on recurring dispatch problems.
Watch It >


Multimedia Thumb

Suspect Steals, Crashes Maryland Ambulance

One killed, others injured in Prince George’s County crash.
Watch It >


Multimedia Thumb

Truck Strikes Pedestrians in Scotland

Six killed in downtown Glasgow.
More >


Multimedia Thumb

Tennessee Trench Rescue

Worker pulled from Roane County worksite.
More >


Multimedia Thumb

Time’s Ebola Firefighters

Doctors, nurses and others saluted for fighting virus.
More >


Multimedia Thumb

LMA MAD Nasal™

Needle-free intranasal drug delivery.
Watch It >


Multimedia Thumb

VividTrac offered by Vivid Medical - EMS Today 2013

VividTrac, affordable high performance video intubation device.
Watch It >


Multimedia Thumb

Field Bridge Xpress ePCR on iPad, Android, Kindle Fire

Sneak peek of customizable run forms & more.
Watch It >


More Product Videos >