FDA to Review Some External Biphasic Defibrillators' Energy Levels


 
 

U.S. Food and Drug Administration | | Wednesday, November 11, 2009


FDA notified healthcare professionals that it is investigating energy levels in external biphasic defibrillators with shocks of less than or equal to 200 Joules (J). FDA received reports of 14 events since 2006 in which a 200 J biphasic defibrillator was ineffective in providing defibrillation/cardioversion therapy to a patient, whereas a subsequent shock from a different 360 J biphasic defibrillator resulted in immediate defibrillation/cardioversion. The majority of events occurred during attempts at cardioversion of atrial fibrillation, but there was at least one instance with defibrillation of a ventricular arrhythmia as well. Analysis of the 14 cases does not suggest the need for any change to current clinical practice, and as FDA continues its evaluation of this situation, providers are encouraged to follow the American Heart Association's guidelines/algorithms for treatment of cardiac arrhythmias, and to follow manufacturers' instructions for using defibrillators.

FDA is seeking additional information in order to interpret the significance of these events, and to determine whether FDA activities are advised. If you suspect a problem with a defibrillator, such as a situation where a patient received shocks from multiple devices, the problem should be reported to the FDA'sMedWatch Safety Information and Adverse Event Reporting Program online (atwww.fda.gov/MedWatch/report.htm,) by phone 800/332-1088, or by returning the postage-paid FDA form 3500 (which may be downloaded from the MedWatch"Download Forms" page) by mail (to address on the pre-addressed form) or fax (800/FDA-0178.)

[11/05/2009 --http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm189259.htm-- FDA]




Connect: Have a thought or feedback about this? Add your comment now
Related Topics: Leadership and Professionalism, Cardiac and Circulation

What's Your Take? Comment Now ...

Featured Careers & Jobs in EMS





 

Get JEMS in Your Inbox

 

Fire EMS Blogs


Blogger Browser

Today's Featured Posts

 

EMS Airway Clinic

Simulation-Based Assessment Facilitates Learning & Enhances Clinical Judgment

Simulation is an educational tool that can be used to develop and refine clinical skills of the student in a controlled environment before they progress to becoming practicing clinicians.
More >

Multimedia Thumb

NYC September 11 Commemoration

Tolling of the bell begins the anniversary ceremony.
Watch It >


Multimedia Thumb

Wrong Fuel Cited in New Mexico Air Ambulance Crash

Plane received jet fuel instead of aviation fuel.
Watch It >


Multimedia Thumb

Wake County EMS PAT

Physical ability test demonstration.
Watch It >


Multimedia Thumb

NYC Sept. 11 Anniversary

View images from the ceremony at Ground Zero.
More >


Multimedia Thumb

13th Anniversary of Sept. 11 Terror Attack

The nation remembers the fallen.
More >


Multimedia Thumb

Building Explosion, Collapse in Paris Suburb

Death toll rises to eight after blast in Rosny-sous-Bois.
More >


Multimedia Thumb

Field Bridge Xpress ePCR on iPad, Android, Kindle Fire

Sneak peek of customizable run forms & more.
Watch It >


Multimedia Thumb

VividTrac offered by Vivid Medical - EMS Today 2013

VividTrac, affordable high performance video intubation device.
Watch It >


Multimedia Thumb

The AmbuBus®, Bus Stretcher Conversion Kit - EMS Today 2013

AmbuBus®, Bus Stretcher all-hazards preparedness & response tool
Watch It >


More Product Videos >