Hands-on Defibrillation

 

 
 
 

Keith Wesley, MD, FACEP | | Sunday, June 15, 2008


Review of:Lloyd MS, Heeke B, Walter PF, et al: "Hands-on defibrillation: an analysis of electrical current flow through rescuers in direct contact with patients during biphasic external defibrillation."Circulation. 117(19):2510-2514, 2008.

The Science

I'll attempt to describe this study without getting too technical for fear of being overwhelmed with e-mails from the techies out there. The researchers who were also the subjects of this study volunteered to maintain contact with a sedated patient who was undergoing elective cardioversion. They wore polyethelene medical gloves and simulated chest compressions. Pre-gelled electrodes similar to those used with AEDs were applied to their chests. The amount of stray current that passed through the "rescuer" was measured during biphasic cardioversion, and defibrillation energies ranged from 100 to 360 joules.

In no cases were shocks perceptible to the rescuer. Peak potential differences between the rescuer's wrist and thigh ranged from 0.28 to 14 V (mean 5.8 +/- 5.8 V). The average leakage current flowing through the rescuer's body for each phase of the shock waveform was 283 +/- 140 uA (range 18.9 to 907 uA). This was below several recommended safety standards for leakage current.

In the study, the researcher concluded, "Rescuers performing chest compressions during biphasic external defibrillation are exposed to low levels of leakage current. The present findings support the feasibility of uninterrupted chest compressions during shock delivery, which may enhance the efficacy of defibrillation and cardiocerebral resuscitation."

The Street

Well I'm not sure I'm ready to go that far, but this will probably be one of the most important papers published this year when we look to the future and the 2010 AHA guidelines. But before we go adopting this radical approach, let's first consider some caveats the authors mention.

What about conduction if the rescuer is not wearing gloves? What's the difference if you're still using the older machines with paddles? We have all seen them arc. The authors don't discuss it, but is there a difference between the conduction with a monophasic machine? I don't know, but I suspect the answer will be that it's far less than we feared.

I continue to encounter services with policies that require them to pull over and come to a complete stop on the side of the road to perform defibrillation. Why? I believe this comes from the fear we had when we introduced this unknown device to EMTs in the early 80's. We said not to touch the patient. Don't defibrillate on a metal grate, wet ground, in the rain, etc, etc. Why? Because we simply didn't know. It's 28 years later, and we haven't had any significant instances of rescuers harmed while providing care.

Perhaps it's time to re-evaluate our operational policies in light of science. Defibrillate safely. Perhaps safely is just slowing down and not taking the turn so sharply, instead of coming to a stop where oncoming traffic can hit you. Perhaps, just perhaps, the AED isn't as dangerous to us as we feared as long as we use it appropriately. And just maybe if this study is repeated in other conditions with other devices, we may be able to keep on pushing while it charges and shocks.




Connect: Have a thought or feedback about this? Add your comment now
Related Topics: Cardiac and Circulation, Provider Wellness and Safety

 
What's Your Take? Comment Now ...

Buyer's Guide Featured Companies

Featured Careers & Jobs in EMS

Get JEMS in Your Inbox

 

Fire EMS Blogs


Blogger Browser

Today's Featured Posts

 

EMS Airway Clinic

Mechanical CPR is Producing Resuscitation Results Beyond Expectations

Discover why clinical studies are finding mechanical CPR just as effective as optimally-performed CPR.
More >

Multimedia Thumb

New York First Responders Take Extraordinary Steps to Help Residents

Snow causes fire and EMS crews to go beyond call of duty.
Watch It >


Multimedia Thumb

Raw Video: Explosion and Fire at California Business

Over a dozen hurt in Santa Paula blaze.
Watch It >


Multimedia Thumb

Four Workers Killed at Texas Chemical Plant

Five employees exposed to methyl mercaptan in La Porte.
Watch It >


Multimedia Thumb

Worker Killed in North Carolina Bridge Collapse

A fatal construction site collapse occurred in Wake County.
Watch It >


Multimedia Thumb

Plane Crashes into Chicago Home

Pilot reported engine trouble after takeoff.
More >


Multimedia Thumb

Switzerland Landslide Kills Two, Injures Four

Slide hits residential building near Lugano.
More >


Multimedia Thumb

World Trade Center Rescue

Technical rescue work is done at nation’s tallest skyscraper.
More >


Multimedia Thumb

LMA MAD Nasal™

Needle-free intranasal drug delivery.
Watch It >


Multimedia Thumb

VividTrac offered by Vivid Medical - EMS Today 2013

VividTrac, affordable high performance video intubation device.
Watch It >


Multimedia Thumb

Field Bridge Xpress ePCR on iPad, Android, Kindle Fire

Sneak peek of customizable run forms & more.
Watch It >


More Product Videos >